Injektron CT 2  Technical Data

Technical Data

Application

  • Fully digital user programmable double-head injector for contrast agent and physiological saline solution (NaCl) in the field of computed tomography

Injector concept

  • Program controlled injection of contrast agent and physiological saline solution (NaCl)
  • Choice between simultaneous and sequential injection of contrast agent and physiological saline solution (NaCl)
  • Heated syringe holder with Easy Loading Syringe (ELS) 200 ml
  • Keep vein open feature
  • Microprocessor controlled double-head injector with electric drive and two independent injection units
  • Keypad with LC-Display on top side of the appliance
  • Wireless Touch Screen remote control
  • Non-stick syringe holder

Design

  • Swivelling injection modules connected to portable housing
  • In operation mode, tilts approx. 20° downwards; in de-aeration mode, tilts vertically upwards
  • Total weight 42 kg

Power supply

  • 230 V/50-60 Hz, 330 VA, optional 115 V/50-60 Hz, 330 VA

Memory

  • 50 programmes can be defined and saved by the user

Filling of injection syringes

  • Automatic filling via menu with volume input or
  • Manual filling with variable speeds
  • Optimised tube systems with automatic valves

Injection parameters

Maximum injection volumes

  • 200 ml (CM)
  • 200 ml (NaCl)

Partial injection volumes

  • 1 - 200 ml, at increments of 1 ml/s

Maximum pressure

  • 21 bar (305 psi)

Delivery rates of the two injection units

  • - 15 ml/s, at increments of 0.1 ml/s

Number of phases

  • 1 to 6

Technical accessories Injektron CT2

Options

  • Adaptor for prefilled syringes
  • Wall or ceiling mounted support
  • CANopen interface  
  • Further options on request

Product information and technical data

 

Product Image

 

Subject to technical alteration