Technical data
|

|
Application
- Fully digital user programmable double-head injector for contrast agent and physiological saline solution (NaCl) in the field of computed tomography
Injector concept
- Program controlled injection of contrast agent and physiological saline solution (NaCl)
- Choice between simultaneous and sequential injection of contrast agent and physiological saline solution (NaCl)
- Heated syringe holder with Easy Loading Syringe (ELS) 200 ml
- Keep vein open feature
- Microprocessor controlled double-head injector with electric drive and two independent injection units
- Keypad with LC-Display on top side of the appliance
- Wireless Touch Screen remote control
- Non-stick syringe holder
Design
- Swivelling injection modules connected to portable housing
- In operation mode, tilts approx. 20° downwards; in de-aeration mode, tilts vertically upwards
- Total weight 42 kg
Power supply
- 230 V/50-60 Hz, 330 VA, optional 115 V/50-60 Hz, 330 VA
Memory
- 50 programmes can be defined and saved by the user
Filling of injection syringes
- Automatic filling via menu with volume input or
- Manual filling with variable speeds
- Optimised tube systems with automatic valves
Injection parameters
Maximum injection volumes
200 ml (NaCl)
Partial injection volumes
- 1 - 200 ml, at increments of 1 ml/s
Maximum pressure
Delivery rates of the two injection units
- - 15 ml/s, at increments of 0.1 ml/s
Number of phases
Options
- Adaptor for prefilled syringes
- Wall or ceiling mounted support
- CANopen interface
Further options on request
Subject to technical alteration

|
© MEDTRON AG
|
|